Use only the B-oil offered by VACUUBRAND. It is characterized by very low vapor pressure, high chemical resistance, and stable behavior toward oxidizing agents as well as acidic and basic vapors. After solvent-intensive processes, the pump should continue running for about 30 minutes with gas ballast open and, if necessary, inlet open to expel residual condensates. The telescopic design considerably facilitates service work.